35 / 144
6•SET-UP
Art.-Nr.: 80116-271
04.06
V 1.2.x Loudsp. test:
Unit checks the loudspeaker function. Three different tones must
be heard.
V 1.3.x Error list IIF/NE:
Stores all errors detected and signaled by the IIF (instrument
interface) and the NE (Nessy2).
V 1.3.x Hardware TP:
Branching to the hardware test programs.
V 1.3.x Upgrade list:
Indicates which upgrades have been installed.
V 1.4.x Enable Kali (only relevant for VIO 200 D):
Makes it possible to increase the HF power limitation for SWIFT
COAG to 150 W. When switching off, the unit resets the increase
back to the standard power limitation of 120 W automatically.
1. Test programs not explained here are not relevant for the service technicians.
2. "safe config." saves the receptacle configuration of the unit detected by the system. The receptacle configuration must be saved by the
service technician after each software update and each time the unit is upgraded or converted. For this purpose compare the receptacle
configuration indicated on the "Version list" with the physical configuration on the unit. If they agree, save the receptacle configuration with
"safe config."