9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
Testing and measuring equipment
IMPORTANT! The following list contains the testing and measuring equipment
recommended by ERBE for servicing. Where ERBE article num
-
bers are specified, only original ERBE testing and measuring
equipment should be used.
ERBE
Art. No
Description
PC/laptop WIN 98 or higher
Oscilloscope, 100 MHz or higher (recommended: Tektronix TDS 1012)
Safety tester (with insulation testing >500 V DC)
HF power meter (recommended: Metron QA-ES)
High Voltage Differential Probe
(recommended: TESTTEC TT-SI 9010, Tektronix P5210 oder
Sapphire SI-9010)
29140-211 VIO HF Adjustment Tool (software, only for internal use by ERBE)
20140-002 VIO Support Hardware (software, only for internal use by ERBE)
20188-100 Single-pedal footswitch
20189-101 Dual-pedal footswith with ReMode
20190-115 VIO ReMode Electrode handle (only if there is an MF receptacle)
20190-045 Electrode handle ICC/ACC
20192-127 Patient cable AE
20192-110 Patient cable AE, international
20196-045 Bipolar cable
20196-053 Bipolar cable, international
20100-034 Adapter cable bipolar
20194-070 Patient cable NE
20194-075 Patient cable NE, international
20100-033 Adapter cable NE
20100-035 Cable LF – leakage current
20100-038 Cable LF – leakage current, international
20100-152 BiClamp measuring cable (only if there is an MF receptacle)
20100-174 Test cable for bipolar resection (only if there is an MF receptacle)
20100-101 VIO Testbox Symmetry/Resistance (NE assymmetry/critical resistance)
20100-102 VIO Testbox Auto Start/Auto Stop (bipolar start/stop)