9 • Troubleshooting
84 / 144
Art.-Nr.: 80116-271
04.06
C A0 Temperature sensor at
CPU + sensors defective.
Check sensor (display in
TP generator or on CAN
bus).
C A1 EEPROM calibration values
incorrect.
Carry out calibration.
C C0 Relay error in APC during
self-check.
Check relay in APC.
C F0 Trial activation with
Pmax==0 (USA) -> no HF.
Information.
Recognizing module: C = CPU + Sensors
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy