To adjust time, press [Print/M+], then input a new value and press [Print/M+] to save.
F6
System (F7~F336)
Initialization
** No
YES
Press [Unit] to shift between NO and YES and press [Print/M+] to confirm.
If YES is selected, instrument displays SURE?”. At this point: -
a. Press [1] to confirm and instrument displays Done when initialization is completed, or
b. Press any key to quit.
F7
Auto Power Off
OFF
1
3
** 5
10
20
Other
Notes: -
Off = Auto power off disabled.
1 ~ 20 = Auto power off after this idle time (minute).
Other = Set specific auto power off time.
All auto power off settings, except Other, will be disabled when a power adaptor is plugged in.
To input a specific auto power-off time: -
1. select Other and press [Print/M+],
2. Input the specific power-off time in the format of HH:MM:SS and press [Print/M+] to save.
F8
Brightness
Brightness (01 ~ 99)
** Default = 60
Color Ratio (01 ~ 99)
** Default = 50
To change setting, input value and press [Print/M+].
Set brightness (bt) first, then set color ratio (Col). The color ratio is used to generate the preferred yellow color.
When this instrument is powered by the built-in rechargeable battery, backlight turns to minimum when the battery is low or when weight
value remains unchanged for 5 seconds.
F9
Weight Unit
Enable/Disable
kg
(** On/Off)
g
(On/** Off)
lb
(On/** Off)
To enable/disable a specific weight: -
1. Set first weight unit kg. Then repeat the below procedures for g and then lb until instrument returns to F9.
2. Press [Unit] or [MR] until the preferred parameter for the weight unit is displayed.
3. Press [Print/M+] to save and go to the next weight unit.