Notes: -
• Set Filter Strength first and then AD Conversion Speed.
• 9 filter strength parameters (1 ~ 9) are available: -
o 1 (strongest filter) for bad working environment where vibration, wind flow etc. affect stable reading,
o 5 for normal environment,
o 9 (least filter) for good working environment where wind and vibration do not affect the stable reading.
• 4 AD conversion speed parameters (15 ~ 120 time per second) are available: -
o 15 times per second (slow speed). Good for high-resolution applications.
o 30 times per second (medium speed). Recommended maximum resolution = 15,000 division
o 60 times per second (high speed). Recommended maximum resolution = 7,500 division
o 120 times per second (ultra-high speed). Recommended maximum resolution = 3,000 division
To set filter strength and AD conversion speed: -
1. Press [Unit] or [MR] to select filter strength. At this point: -
a. Press [Unit] or [MR] to select the preferred filter strength, then [Print/M+] to confirm, or
b. Press [Zero] to quit
2. Instrument displays the AD conversion speed setting page. At this point: -
a. Press [Unit] or [MR] to select the preferred AD conversion speed, then [Print/M+] to confirm, or
b. Press [Zero] to quit. If AD conversion speed is changed, instrument will restart automatically.
To enable and disable a specific Auxiliary Function: -
1. Press [Unit] or [MR] until the preferred parameter is displayed.
2. Press [Print/M+] to save and go to the next Auxiliary working mode.