Data Transfer
ESC9 Operates the MENU key
G0 or
ES or
EN or
ESC?
Trigger a measurement; the measurement reading is saved in the
instrument.
Response of the instrument: Measurement reading
GAA Requests the number of the current Application.
GAN0
...
GAN99
Requests the designation of the Application with the number 0
(... 99). An empty string is output if no designation has been as-
signed.
Assign a designation using the software Fischer DataCenter or
the command “SAN”.
Example: After receiving the command “GAN2”, the instrument
outputs the name of the second Application.
GBN0
...
GBN999
Requests the designation of the measurement block with the num-
ber 1 (... 1000). An empty string is output if no designation has
been assigned. Assign a designation using the software Fischer-
DataCenter or the command “SBN”.
Example: After receiving the command “GBN2”, the instrument
outputs the designation of the third block of the current Applica-
tion.
GG Requests the block size of the current Application, i.e., the num-
ber of readings that are to be combined during automatic block
creation. “0” is output if automatic block creation is disabled.
Enables automatic block creation and sets the block size of the
current Application: 5.6.3 ‘Automatic Block Size and Block
Creation’, beginning on Page 49.
GGS Requests the group separator mode.
The instrument outputs “1” if group separator is enabled (i.e., a
group separator is transmitted between each measurement block
during the output via the interface port). “0” is output if the
group separator is disabled.
Setting the group separator mode: Command “SGS1” or
9.4.3 ‘Transferring Data With a Group Separator’, begin-
ning on Page 135.
GMX Requests the measuring mode. The instrument outputs “1” if the
matrix mode is enabled. A “0” output indicates that the standard
measuring mode is enabled.
Enabling the matrix measuring mode: 10.6.3 ‘Matrix Mode’,
beginning on Page 169.
Command Function