ProSim 8
Users Manual
88
Dynamic Waveforms
Types (default pressures) ................................... Arterial (120/80)
Radial artery (120/80)
Left ventricle (120/00)
Right ventricle (25/00)
Pulmonary artery (25/10)
Pulmonary-artery wedge (10/2)
Right atrium (central venous or CVP) (15/10)
Pressure Variability ............................................. Systolic and diastolic pressures are independently variable in 1 mmHg
steps.
Swan-Ganz Sequence ........................................... Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary
artery wedge (PAW)
Cardiac Catheterization
Chambers ........................................................... Aortic, Pulmonary valve, and Mitral valve
Respiration Artifact
Arterial, radial artery, and left ventricle ............... 5 % to 10 % multiplication
Other ................................................................... 5 mmHg or 10 mmHg
BP Output .............................................................. Circular DIN 5-Pin
Power-On Default .................................................. 0 mmHg
Respiration
Rate ........................................................................ 0 BrPM (OFF), 10 BrPM to 150 BrPM in 1 BrPM steps
Waves ..................................................................... Normal or ventilated
Ratio (inspiration:expiration)
Normal ................................................................ 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated ............................................................ 1:1
Impedance Variations (Δ Ω) ................................. 0.00 Ω to 1.00 Ω in 0.05 Ω steps and 1.00 Ω to 5.00 Ω in 0.25 Ω steps
Delta Accuracy ...................................................... ±(5 % of setting + 0.1 Ω)
Baseline ................................................................. 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III
Baseline Accuracy ................................................ ±5 %
Respiration Lead ................................................... LA or LL (default)
Apnea Selection .................................................... 12 seconds, 22 seconds, or 32 seconds (one-time events), or
continuous (Apnea ON = respiration OFF)
Power-On Default .................................................. 20 BrPM, delta 1.0 Ω
Temperature
Temperature .......................................................... 30.0 °C to 42.0 °C in 0.5 °C steps
Accuracy ................................................................ ±0.4 °C
Compatibility ......................................................... Yellow Springs, Inc. (YSI) Series 400 and 700
Output .................................................................... Circular DIN 4-pin
Cardiac Output
Catheter Type ........................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient .......................................... 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature ............................................... 36 °C (98.6 °F) to 38 °C (100.4 °F) ±0.2 °C in 1 °C steps
Injectate Volume ................................................... 10 cc
Injectate Temperature........................................... 0 °C or 24 °C
Cardiac Output ...................................................... 2.5 liters per minute, 5 liters per minute, 10 liters per minute ±7.5 %
Faulty-Injectate Curve .......................................... Waveform for simulation available
Left-to-Right-Shunt Curve .................................... Waveform for simulation available
Calibrated Pulse .................................................... 1.5 °C for 1 second
Connector .............................................................. Circular DIN 7 pin
Power-On Default ................................................. 5 liters per minute, 0 °C injectate, 37 °C blood temperature