Vital Signs Simulator
Detailed Specifications
87
R-wave Detection
Waveform............................................................ Triangular pulse
Rate .................................................................... 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Width ................................................................... 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width Accuracy ................................................... ±(1 % of setting + 1 ms)
QRS Detection
Widths ................................................................. 8 ms to 20 ms in 2 ms steps and 20 ms to 200 ms in 10 ms steps
Width Accuracy ................................................... ±(1 % of setting + 1 ms)
Rate .................................................................... 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
R-Wave up slope ................................................ 0.875 amplitude, 0.4375 x width
R-Wave down slope ............................................ Full amplitude, 0.5 x width
S-Wave up slope ................................................ 0.125 amplitude, 0.0625 x width
Tall T-Wave Rejection
Waveform
QT Interval ...................................................... 350 ms
T-Wave width .................................................. 180 ms
T-Wave Shape ................................................ ½ sinewave
Amplitude ............................................................ 0 % to 150 % reference lead amplitude in 10 % steps
Rate .................................................................... 80 BPM
Rate Accuracy ....................................................... ±1 % of setting
Amplitude Accuracy ............................................. ±(2 % of setting + 0.05 mV)
ECG Artifact
Type ........................................................................ 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size ......................................................................... 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
Lead Select ............................................................ All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal / Maternal-ECG
Fetal Heart Rate (Fixed) ........................................ 60 BPM to 240 BPM in 1 BPM steps
Fetal Heart Rate (IUP) ........................................... 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ....................... Early deceleration, late deceleration, and acceleration
Wave Duration ....................................................... 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg
and returning to 0 mmHg
IUP Period .............................................................. 2 minutes, 3 minutes, or 5 minutes; and manual
Default Settings ..................................................... FHR 140 BPM, early deceleration wave, manual
Invasive Blood Pressure
Channels ................................................................ 2, each independently settable with identical parameters and are
individually electrically isolated from all other signals
Input/output Impedance ....................................... 300 Ω ±10 %
Exciter Input Range .............................................. 2.0 V to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5 μV/V/mmHg (default) or 40 μV/V/mmHg
Pressure Accuracy ............................................... ±(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation
only
Static Pressure ...................................................... -10 mmHg to +300 mmHg in 1 mmHg steps
Pressure Units ....................................................... mmHg or Kpa