ProSim 8
Users Manual
90
Transmission (Ratio of detector current to LED current, expressed in parts per million (ppm))
Range ................................................................. 0 ppm to 300.00 ppm
Resolution ........................................................... 0.01 ppm
Accuracy ............................................................. +50 %/-30 % for compatible monitors, unspecified for others. Selected
by finger size and color: Dark, thick finger, medium finger, light, thin
finger, neonatal foot.
Pulse Amplitude
Range ................................................................. 0 % to 20.00 %
Resolution ........................................................... 0.01 %
Artifact
Respiration
Range ............................................................. 0 % to 5 % of transmission
Resolution ....................................................... 1 %
Rate ................................................................ all ProSim respiration simulation settings
Ambient Light
Range ............................................................. 0X to 5X transmitted light
Resolution ....................................................... 1X
Frequency ....................................................... DC, 50 Hz, 60 Hz, and 1 kHz to 10 kHz in 1 kHz steps
Masimo Rainbow Technology ............................. Masimo Rainbow technology with an optional adapter supplied by
Masimo that allows the ProSim two wavelength to test the Rainbow
multiple wavelength system
Compatible Manufacturer Products
With manufacturer R-curve ................................. Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke Biomedical R-curve ........................... Mindray, GE-Ohmeda, Philips/HP, and BCI
Pre-Defined Simulations
Normal
Hypertensive
Hypotensive
Tachycardic
Bradycardic
Ventricular Fibrillation
Asystole
Autosequences (default)
Monitor testing sequence
Medical training sequence
Oximeter testing sequence
Cardiac failure sequence
Arrhythmia sequence
Exercise sequence
Respiration sequence
Performance Wave Test
IBP testing sequence
Temperature sequence