ProSim 8
Users Manual
20
The Run of PVCs arrhythmia has a number variable to set the number of PVCs
in a run. To change the number of PVCs in a run:
1. Push or to highlight the Number value.
2. Push .
3. Push or to highlight 2 PVCs, 5 PVCs, or 11 PVCs.
4. Push to set the arrhythmia number and go back to the ECG: Ventricular
screen.
Conduction Arrhythmias
Conduction arrhythmias are caused when conduction of electrical impulses are
stopped or blocked from their usual pathways around the heart. To simulate a
conduction arrhythmia:
1. From the ECG screen, push or to highlight the Wave Group value.
2. Push .
3. Push or to highlight Conduction in the wave group list.
4. Push .
5. Push or to highlight the Waveform value.
6. Push .
7. Push or to highlight a waveform in the list.
8. Push to set the arrhythmia waveform and go back to the ECG:
Conduction screen.
You can adjust the amplitude of all conduction arrhythmia waveforms. See Set
the ECG Parameters. You can also add ECG signal artifacts to each arrhythmia
signal. See Simulate an ECG Artifact.
Pacemaker Waveforms
The Product can simulate ECG waveforms with a number of artificial-pacemaker
conditions. Table 11 shows the simulations in the waveform list of the Product
with a description of each.
Table 11. Pacemaker Waveforms
Pacemaker Waveform Label Waveform Description
Atrial 80 BPM
Atrial pacemaker wave at 80 BPM, with a pacer pulse at the start of each
P wave.
Asynchronous 75 BPM
Asynchronous pacemaker wave with continuous ventricular-paced beats
(75 BPM) and no P waves
Demand Freq Sinus
A “demand” pacemaker wave with frequent sinus beats (forty normal
beats followed by twenty ventricular-paced beats, repeated)
Demand Occ Sinus
A “demand” pacemaker wave with occasional sinus beats (twenty normal
beats followed by forty ventricular-paced beats, repeated)
Atr-Vent Sequential
AV-sequential-pacemaker wave with continuous paced beats, each with
an atrial pulse and a P wave followed by a ventricular-paced pulse and
QRS response
Non-Capture
Ventricular-paced beats, where one out of ten beats has no heart
response.
Non-Function Continuous pacer pulses at 75 BPM with no heart response.