Vital Signs Simulator
ECG Function
21
To select a pacemaker waveform simulation:
1. From the ECG screen, push or to highlight the Wave Group value.
2. Push .
3. Push or to highlight TV Paced in the wave group list.
4. Push .
5. Push or to highlight the Waveform value.
6. Push .
7. Push or to highlight a waveform in the list.
8. Push to set the arrhythmia waveform and go back to the ECG: TV
Paced screen.
You can adjust the amplitude of all TV Paced waveforms. See Set the ECG
Parameters. You can also add ECG signal artifacts to each arrhythmia signal.
See Simulate an ECG Artifact.
TV-Paced waveforms include the applicable pacer pulses. You can change the
pacer pulses through the front panel. You can change all the ventricle pacemaker
variables except for the Atrial 80 BPM waveform. The Atrial 80 BPM has an atrial
pacemaker value. The Atr-Vent Sequential waveform has an atrial pacemaker
value and a ventricle pacemaker value. To change the pacemaker values, push
the Atrial Settings or Vent. Settings softkeys.
You can change the amplitude, width, and polarity of the pacemaker signal. To
change the amplitude:
1. Push or to highlight the Amplitude value.
2. Push .
3. Push or to highlight an Amplitude value in the list. Below is a list of the
pacemaker amplitude values.
0 mV 16 mV
2 mV 18 mV
4 mV 20 mV
6 mV 50 mV
8 mV 100 mV
10 mV 200 mV
12 mV 500 mV
14 mV 700 mV
4. Push to set the amplitude and go back to the TV Paced Settings
screen.
To change the width value:
1. Push or to highlight the Width value.
2. Push .
3. Push or to highlight 0.1, 0.2, 0.5, 1.0, 1.5, or 2.0 mV in the list.
4. Push to set the width and go back to the TV Paced Settings screen.