Vital Signs Simulator
Oximeter SpO2 Optical Emitter and Detector
63
To change the transmission value:
1. From the SpO2 screen, push or to highlight the Transmission value.
2. Push .
3. Push or to highlight Dark, Thick Finger, Medium Finger, Light, Thin
Finger, and Neonatal Foot.
4. Push to set that transmission value and go back to the SpO2 screen.
You can change the simulated heart rate of the Product from the SpO2 screen.
To change the heart rate:
1. Push or to highlight the Heart Rate value.
2. Push .
3. Push or to adjust the heart rate. Each push of a key moves the heart
rate one beat in the direction of the key pushed. The range is 10 BPM to
300 BPM.
4. Push to go back to the SpO2 screen.
To change SpO2 respiration:
1. From the SpO2 screen, push the or to highlight the Respiration value.
2. Push .
3. Push or to highlight 1%, 2%, 3%, 4%, 5% or off.
4. Push to set the respiration and go back to the SpO2 screen.
You can test SpO2 under different ambient light conditions. To change the
ambient light value:
1. From the SpO2 screen, push the or to highlight the Ambient Light
value.
2. Push .
3. Push or to highlight an ambient light value. Below is the list of ambient
light values in the Product.
Sunlight Indoor – 4 kHz
Indoor – DC Indoor – 5 kHz
Indoor – 50 Hz Indoor – 6 kHz
Indoor – 60 Hz Indoor – 7 kHz
Indoor – 1 kHz Indoor – 8 kHz
Indoor – 2 kHz Indoor – 9 kHz
Indoor – 3 kHz Indoor – 10 kHz
4. Push to go back to the SpO2 screen with the new ambient light value.