L9-1.2
Tubing
length
(mm)
ConnectorTubingTubing
label
Connection
between...
100For ÄKTA avant
25: Fingertight
connector, 1/16"
For ÄKTA avant
25: PEEK, o.d.
1/16", i.d. 0.50 mm
5L1
1
V9-Inj or V9H-Inj
and L9-1.2
For ÄKTA avant
150: Fingertight
connector, 1/16"
For ÄKTA avant
150 : PEEK, o.d.
1/16", i.d. 1 mm
100For ÄKTA avant
25: Fingertight
connector, 1/16"
For ÄKTA avant
25: PEEK, o.d.
1/16", i.d. 0.50 mm
5L2
1
L9-1.2 and V9-C or
V9H-C
For ÄKTA avant
150: Fingertight
connector, 1/16"
For ÄKTA avant
150: PEEK, o.d.
1/16", i.d. 1 mm
1
Cut the spare piece of tubing delivered with the system to correct length.
System properties
Follow the instruction to update the system properties.
ActionStep
Open the system properties Edit dialog box.1
In the Component types list, click Monitors and sensors.2
Select the External air sensor (L9) check box in the Component selection
list.
3
In the Property list, click Before sample inlet or After injection valve depend-
ing on where the air sensor is placed.
Note:
The available properties are the same, regardless of which type of air sensor
that is used, L9-1.2 or L9-1.5.
4
Click OK.5
122 ÄKTA avant User Manual 29035184 AE
4 Optional instrument modules
4.6 External air sensors