7.6.5 Clean Fraction collector F9-R
Maintenance interval
Clean the Fraction collector when required, for example in case of liquid spill.
Required material
The following materials are required:
•
Water or 20% ethanol
•
Cloth
Clean the instrument
Follow the instructions below to clean the instrument externally.
ActionStep
Check that no run is in progress.1
Switch off the instrument.2
Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning
agent or 20% ethanol. Wipe off any excess.
3
Let the Fraction collector F9-R dry completely before restart.4
Clean DropSync sensor
Clean the drop sensor photocell located above the tube sensor (see Front view illustration,
on page 131) with a damp cloth.
304 ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.6 Maintenance when required
7.6.5 Clean Fraction collector F9-R