8.3 Troubleshooting: Monitors
In this section
UV monitor and UV detector
Conductivity monitor
pH monitor and pH valve
Pressure monitors
UV monitor U9-M and UV
detector unit
Possible cause and actionProblem
Communication problem
Contact Service.
The cable between the UV module and the ICU is
not connected
Remove the UV module and make sure that the cable
is connected.
Wrong Node ID
Check the module´s Node ID. If necessary, change the
Node ID. See Section 9.15 Check and change the Node
ID of a module, on page 515.
The UV module is not
found by the instru-
ment
ÄKTA avant User Manual 29035184 AE 363
8 Troubleshooting
8.3 Troubleshooting: Monitors