9.7.6 System settings - I/O-box
The following table describes the I/O-box related system settings available for ÄKTA
avant.
DescriptionInstruction name
Sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Digital out X
Filters the noise in the analog signal in port number X.Noise reduction analog
in X
Enables or disables the alarm for the analog signal in
port number X. When enabled, it sets the alarm limits
for the analog signal. If the alarm is enabled and the
analog signal falls outside the set limits, an alarm will
be triggered and the method will be paused.
Alarm analog in X
Enables or disables the alarm for the signal in digital
port number X. The alarm can be triggered by either of
the signal values, 0 or 1. If the alarm is enabled and the
condition set in Value occurs, an alarm will be triggered
and the method will be paused.
Alarm digital in X
Enables the user to send one of the pre-defined signals
(UV signal, conductivity, temperature, pH or concentra-
tion of eluent B) to the analog out port number X, and
also to set the range of that signal.
Configure analog out X