7.6.12 Clean the Conductivity flow cell
Maintenance interval
Clean the Conductivity flow cell when required.
Required material
The following materials are required:
Luer connector
Waste container
Syringe, 25 ml
1 M NaOH
Distilled water
Instruction
Follow the instruction to clean the flow cell of the Conductivity monitor. The same proce-
dure also applies to the second Conductivity monitor.
ActionStep
Disconnect the Fingertight connector and the piece of tubing from the top
of the Conductivity monitor, and attach a Luer connector.
1
Disconnect the piece of tubing from the bottom of the Conductivity monitor,
and connect a piece of waste tubing to the Conductivity monitor. Place the
waste tubing in a waste container.
2
Disconnect the Conductivity monitor from the rails, or slide the Conductivity
monitor to the right on the rails.
3
318 ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.6 Maintenance when required
7.6.12 Clean the Conductivity flow cell