9.2 Module specifications
Introduction
This section specifies the operating data of the components in ÄKTA avant. For general
data for the system see Section 9.1 System specifications, on page 422.
System pumps
DataParameter
Piston pump, metering typePump type
ÄKTA avant 25: 0.001 to 25 ml/minFlow rate range (normal range)
ÄKTA avant 150: 0.01 to 150 ml/min
ÄKTA avant 25: 0.01 to 50 ml/minFlow rate range (column packing flow)
ÄKTA avant 150: 0.01 to 300 ml/min
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)Pressure range
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: 0.35 to 10 cPViscosity range
ÄKTA avant 150: 0.35 to 5 cP
ÄKTA avant 25:Flow rate
specifications
Accuracy: ± 1.2%
Precision: RSD < 0.5%
(Conditions: 0.25 to 25 ml/min, < 3 MPa, 0.8 to 2 cP)
ÄKTA avant 150:
Accuracy: ± 1.5%
RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)
424 ÄKTA avant User Manual 29035184 AE