Sample pump
DataParameter
Piston pump, metering typePump type
ÄKTA avant 25: 0.001 to 50 ml/min
ÄKTA avant 150: 0.01 to 150 ml/min
Flow rate range
ÄKTA avant 25:0 to 10 MPa (0 to 100 bar)
ÄKTA avant 150: 0 to 5) MPa (0 to 50 bar)
Pressure range
0.7 to 10 cPViscosity range
ÄKTA avant 25:Flow rate
specifications
•
Accuracy: ± 2%
•
RSD < 0.5%
(Conditions: 0.25 to 50 ml/min, < 3 MPa, 0.8 to 3 cP)
ÄKTA avant 150:
•
Accuracy: ± 2%
•
RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)
Mixer
DataParameter
Chamber with magnetic stirrerMixing principle
ÄKTA avant 25: 0.6, 1.4 (default) or 5 mlMixer volume
ÄKTA avant 150: 1.4, 5 (default) or 15 ml
ÄKTA avant User Manual 29035184 AE 425
9 Reference information
9.2 Module specifications