ActionStep
Re-install the tubing holder in the delivery arm.5
Connect the tubing from the fraction collector to port Out 10 on the Outlet
valve.
6
Adjust the delay volume setting in UNICORN to the volume of the tubing, see
Set the delay volume in UNICORN, on page 506 for more details.
7
System properties
Follow the instruction to update the system properties.
ActionStep
Open the system properties Edit dialog box.1
In the Component types list, click Fraction collectors.2
Select the Fraction collector 2 (F9-R) check box in the Component selection
list. Then click OK.
3
The Fraction collector check box must be selected for the Fraction collector
2 (F9-R) check box to be available.
Note:
the Fraction collector 2 (F9-R) check box is not available if the second or third
Outlet valve is installed.
Note:
ÄKTA avant User Manual 29035184 AE 135
4 Optional instrument modules
4.9 Fraction collector F9-R