Choose and start a method
The following instruction describes how to open a method and start a run.
ActionStep
Open the System Control module and click the Open Method Navigator
button.
1
Result: The Method Navigator pane opens.
Select the method to run, and click the Run button.2
Result: The Start Protocol dialog box opens.
Step through the displayed pages in the Start Protocol, add requested input
and make appropriate changes if necessary. Click Next.
3
Click Start on the last page of the Start Protocol.4
Result:
If column logging was chosen during installation of UNICORN and a
column type was selected when the method was created, the Select
Columns dialog box opens. For further information on column handling,
please refer to UNICORN Method Manual and UNICORN System Control
Manual.
If column logging was not chosen during installation of UNICORN and/or
no column type was selected when the method was created, the run
starts directly.
ÄKTA avant User Manual 29035184 AE 231
5 Operation
5.9 Create a method and perform a run
5.9.2 Perform a run