Clean the manual injection port
of the injection valve
Follow the instructions to manually clean the manual load position of the Injection valve.
ActionStep
In the System Control module, on the Manual menu, click Execute Manual
Instructions.
1
Result: The Manual instructions dialog opens.
In the Manual instructions dialog, select Flowpath:Injection valve, and click
Manual Load. Click Execute.
2
Connect a suitable sample loop to Injection valve ports LoopF (fill) and LoopE
(empty).
Note:
Do not use a Superloop when cleaning the Injection valve.
3
Connect tubing to Injection valve port W1, and direct this tubing to a waste
container.
4
294 ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.6 Maintenance when required
7.6.2 Perform System CIP