ActionStep
In the Calibration dialog, in the Monitor to calibrate list, select the pressure
monitor to calibrate.
2
Click Reset pressure.3
Result: The atmospheric pressure is defined as zero. Date and time of the
most recent calibration, and the current pressure value are displayed.
Reconnect the tubing to the pressure monitor.4
324 ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.7 Calibration procedures
7.7.2 Calibrate the pressure monitors