Possible cause and actionProblem
The air intake on the rear or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
The internal tempera-
ture of the UV monitor
is too high
No UV flow cell is attached
Attach UV cell. See Section 7.8.4 Replace the UV moni-
tor U9-M flow cell, on page 342.
UV flow cell is not correctly installed.
Verify that the UV cell is correctly installed. See Sec-
tion 7.8.4 Replace the UV monitor U9-M flow cell, on
page 342
The UV flow cell is broken
Replace the cell. See Section 7.8.4 Replace the UV
monitor U9-M flow cell, on page 342.
UV cell path length un-
readable
ÄKTA avant User Manual 29035184 AE 365
8 Troubleshooting
8.3 Troubleshooting: Monitors