Possible cause and actionProblem
•
The air intake on the rear or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
•
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
•
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
The internal Mixer
temperature is too
high
Superloop
Possible cause and actionProblem
•
The Superloop is filled to the max. Pressure is not
released anywhere.
Manually turn Injection valve to Manual load position.
Overpressure during
filling
Cabinet
Possible cause and actionProblem
•
The lock/unlock knob is in the unlock position
Turn the lock/unlock knob to the lock position.
•
The rubber behind the lock/unlock knob is unclean
Clean the swirl foot.
•
The rubber behind the lock/unlock knob is worn out
The rubber needs to be changed. Call service.
The instrument is not
locked on the chosen
swirl foot position
•
Wrong procedure
Pull the second section of the foldable door upwards
before opening it.
The second section of
the foldable door can-
not be opened
406 ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.7 Troubleshooting: Other components