Possible cause and actionProblem
Cables incorrectly connected
Make sure that the cables are connected correctly,
earth to earth etc.
To isolate the problem:
External digital equip-
ment does not respond
to set changes in Digi-
tal Out
1
Connect Digital Out (e.g. pin 6) to Digital In (pin 1,2,3,
or 4).
2
Change the Digital Out 1 signal between ”1” and ”0”.
Verify that Digital out follows.
3
If not: contact Service.
If it does: the problem probably lies within the connect-
ed equipment.
Auto-zero on the wrong level
Reset Auto-zero.
Digital Out connection error
Make sure that Digital Out is connected correctly, earth
to earth etc.
Analog In does not
measure expected
voltage
Too long cable between the external equipment and
E9
Use as short cable as possible. Use shielded cable.
Connect the cable shield to the D-sub connector’s
shield.
Noisy analog signal
Analog In is not calibrated
Contact Service.
Analog In does not
measure expected
voltage
Analog Out is not calibrated
Contact Service.
Analog Out does not
generate expected
voltage
Instrument display
Possible cause and actionProblem
Display is broken
Generate a System error report and contact Service.
Absence of or incorrect
colors