EasyManua.ls Logo

GE AKTA avant - Page 414

GE AKTA avant
538 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Possible cause and actionScenario
Cause: Two or several modules have the same Node ID.Warning messages in
UNICORN: (Warning)
Two instrument mod-
ules have the same
Node ID.
Action:
1
Switch off the instrument.
2
Check the Node ID for all modules, see Section 9.15
Check and change the Node ID of a module, on
page 515.
3
Switch on the instrument.
4
In the displayed dialog box in UNICORN, select the
option Restart the system only and click OK.
Cause: One module has an incorrect Node ID.Warning messages in
UNICORN: (Warning)
Gate (12): Internal in-
strument error.
Action:
1
Switch off the instrument.
2
Check the Node ID for all modules, see Section 9.15
Check and change the Node ID of a module, on
page 515.
3
Switch on the instrument.
4
In the displayed dialog box in UNICORN, select the
option Restart the system only and click OK.
414 ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.8 Troubleshooting: Instrument communication