Number of inlets
DataParameter
7, expandable to 14Inlet A
7, expandable to 14Inlet B
7, expandable to 14Sample inlet
4Quaternary inlet
Pressure monitors
DataParameter
4Number of sensors
The System pressure monitor is located after the System
pumps.
The Pre-column pressure monitor and the Post-column
pressure monitor are integrated in the Column valve V9-
C or V9H-C.
The Sample pressure monitor is located after the Sample
pump.
Placement of sensors
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)Range
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: ± 0.02 MPa or ± 2% whichever is greaterAccuracy
ÄKTA avant 150: ± 0.015 MPa or ± 1.5% whichever is greater
UV monitors
DescriptionItem
Up to 2Number of monitors
U9-M: 190 to 700 nm in steps of 1 nm, up to 3 wavelengthsWavelength range
U9-L: 280 nm
-6 to 6 AUAbsorbance range
ÄKTA avant User Manual 29035184 AE 427