EasyManua.ls Logo

GE AKTA avant - Page 457

GE AKTA avant
538 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DescriptionPhase Name
Unit selection for Method base and Flow rate
Monitor settings:
-
pH monitor
-
Air sensor alarm settings
-
UV monitor
Note:
The first wavelength of U9-M and the fixed wavelength for
U9-L or U9-L 2nd is always turned on. The second and third
wavelengths for UV monitor U9-M can be set on or off.
Settings for Column Logbook
Start Protocol
Result name and location
Note:
Some of these options may not be required by certain methods.
Note:
The Method Settings phase differs for systems that have the Method
settings phase and the User defined phase as the only available
predefined phases.
Cleans the column after purification runs by rinsing the column
with a cleaning solution to remove non-specifically bound proteins.
By adding steps, several cleaning solutions can be used sequen-
tially.
Column CIP
Tests the efficiency of a packed column in terms of height equiva-
lent to a theoretical plate (HETP), and the peak asymmetry factor
(A
s
).
Column Perfor-
mance Test
Prepares the column before use by removing the storage solution
and equilibrating the column. By adding steps, several preparation
solutions can be used sequentially.
Column Prepa-
ration
Washes out unbound sample after sample application or removes
strongly bound proteins after elution.
Column Wash
Elutes the sample from the column. Defines parameters for the
elution and fractionation settings.
Elution
ÄKTA avant User Manual 29035184 AE 457
9 Reference information
9.6 Predefined methods and phases

Table of Contents

Other manuals for GE AKTA avant

Related product manuals