EasyManua.ls Logo

GE AKTA avant - Page 480

GE AKTA avant
538 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DescriptionInstruction name
Turns the numbered Versatile valve to the selected po-
sition.
Note:
Four sets of positions are available. In positions 1-3 and
2-4 only a single flow channel can be used. In positions
1-4 & 2-3 and 1-2 & 3-4 the flow can be directed through
two channels simultaneously.
Versatile valve to Versa-
tile valve 4
Turns the extra inlet valve to the selected position. The
extra valve is a basic 8 port valve, without air sensor, to
be used for general applications.
Valve X1
Turns the extra inlet valve to the selected position. The
extra valve is a basic 8 port valve, without air sensor, to
be used for general applications.
Valve X2
Sets an injection mark in the chromatogram at the point
where this instruction is executed.
Note:
The instruction is useful when the sample is loaded onto
the column by the system pump.
Injection mark
480 ÄKTA avant User Manual 29035184 AE
9 Reference information
9.8 Using Manual instructions
9.8.2 Manual instructions - Flow path

Table of Contents