DescriptionInstruction name
Enables or disables the alarm for the UV signal from UV
monitor U9-L, 2nd. When enabled, it sets the alarm
limits for the UV signal from U9-L, 2nd. When the UV
signal falls outside the set limits, an alarm will be trig-
gered and the method will be paused.
Alarm UV 2nd
Enables or disables the conductivity alarm. When en-
abled, it sets the alarm limits for the conductivity signal.
When the conductivity falls outside the set limits, an
alarm will be triggered and the method will be paused.
Alarm conductivity
Enables or disables the pH alarm. When enabled, it sets
the alarm limits for the pH signal. When the pH falls
outside the set limits, an alarm will be triggered and the
method will be paused.
Alarm pH
Enables or disables the air sensor alarm for the built-in
air sensor at inlet A. If the alarm is enabled and air is
detected, an alarm will be triggered and the method
will be paused.
Alarm inlet A air sensor
Enables or disables the air sensor alarm for the built-in
air sensor at inlet B. If the alarm is enabled and air is
detected, an alarm will be triggered and the method
will be paused.
Alarm inlet B air sensor
Enables or disables the air sensor alarm for the built-in
air sensor at the sample inlet. If the alarm is enabled
and air is detected, an alarm will be triggered and the
method will be paused.
Alarm sample inlet air
sensor
Enables or disables the alarm for the optional air sensor
number X. If the alarm is enabled and air is detected,
an alarm will be triggered and the method will be
paused.
Alarm external air sensor
X
Enables or disables the alarm for the analog signal in
port number X. When enabled, it sets the alarm limits
for the analog signal. If the alarm is enabled and the
analog signal falls outside the set limits, an alarm will
be triggered and the method will be paused.
Alarm analog in X
490 ÄKTA avant User Manual 29035184 AE
9 Reference information
9.8 Using Manual instructions
9.8.6 Manual instructions - Alarms