EasyManua.ls Logo

GE Hydran M2-X - Appendix A. Technical Specifications; General Specifications; Analytical Performance

GE Hydran M2-X
204 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 145
Hydran* M2-X Instruction Manual
MA-029 Rev. 1.0, 30-Jul-18
Appendix A. Technical Specifications for the Hydran M2-X
A-1. General
GENERAL
Description
Continuous, on-line, intelligent gas-in-oil and moisture
transmitter with optional, external sensors monitoring
Components
Combined dual-function sensor and electronic enclosure
Response
Hydrogen (H
2
), carbon monoxide (CO), acetylene (C
2
H
2
), ethylene
(C
2
H
4
), relative humidity in oil (%RH)
Medium
Mineral, insulating oil for transformers
Application
Transformer monitoring, moisture level measurement (for the
evaluation of dangerous conditions, bubbling temperature and
aging rate) and detection of incipient faults in oil-filled electrical
equipment
ANALYTICAL PERFORMANCE
Principle
• Gas: Vacuum-resistant gas-permeable membrane and
combustible gas detector
• Moisture: Thin-film capacitive sensor
Sampling Method
• Flooded port with 1.5 in NPT male threads
Measurement
• Gas: 25-2,000 ppm (volume/volume, H2 equivalent)*
Range
• Moisture: 0-100% RH
Accuracy
• Gas: ± 10 % of reading ± 25 ppm (Ref.: ASTM D3612)
• Moisture: ± 2 %RH (Ref.: saturated salt solutions)
Repeatability
Limit
• Gas: Better than ± 5 % or ± 5 ppm
• Moisture: 2 %RH
Gas Relative
• H2: 100 % of concentration
Sensitivity
• CO: Typical 15 ± 4 % of concentration
• C2H2: Typical 8 ± 2 % of concentration
• C2H4: Typical 1.5 ± 0.5 % of concentration
Response Time
• 10 minutes sensor response (90 % of step change)

Table of Contents