Figure 3-3:  Example of Unacknowledged Alarm 
Each screen contains: 
•  The name of the parameter responsible for the alarm. 
•  The value of this parameter. 
•  The date and time at which the alarm condition was detected. 
 
The unacknowledged  alarm screens are always displayed in priority.   When an 
alarm  condition  occurs,  the  Main  Display  mode  is  immediately  replaced  by  the 
Unacknowledged Alarms mode. 
An  alarm  is  acknowledged  by  pressing  the  Ackn.  context  key  while  the 
corresponding message is displayed.  When acknowledged, the message is removed.  
The  Hydran  M2-X  returns  to  the  Main  Display  mode  when  all  alarms  have  been 
acknowledged. 
Acknowledged  alarms  that  are  still  active  are  shown  in  a  screen  of  the  Main 
Display  mode  (see  Section  3.1.2.3).    When  an  alarm  condition  disappears,  the 
corresponding message is removed from the Main Display mode screen. 
3.1.2.2  Menu Navigation Mode 
In  the  Menu  Navigation  mode,  the  keypad  is  used  to  access  the  Main  Menu’s 
options, parameters and commands.  The Main Menu gives access to all Hydran M2-
X’s parameters and  commands   To  access  the  Main  Menu (from  the  Main Display 
mode), press the Main Menu context key. 
Each option of the Main Menu is described in detail in Section 3.3 to Section 3.6. 
Note: The Hydran M2-X automatically returns to the Main Display mode if the keypad 
is not used for five minutes. 
3.1.2.3  Main Display Mode 
This mode is displayed automatically if there are no unacknowledged alarms and 
the keypad is unused for more than five minutes.  To select this mode manually from 
the Main Menu, press the Esc key until screen 1 (described below) of the Main Display 
mode is reached. 
In the Main Display mode, the Hydran M2-X displays four screens of messages: