DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3 - 30 Section 3-4 - Completing the setup
16.)Use one of the methods to access/unlock the system:
- Connect a USB Flash Drive to the system.
-Select Password and type the password in the field.
-Select Recovery key and then type in the recovery key in the field.
Secure the password and recovery key (including USB memory)
It is recommended the recovery key is stored in a USB flash drive and printed out.
If both password and recovery key are lost, you will not be able to access archived patient data (images
and measurements included) nor store new patient data on this system.
The only way to recover the system to allow storing patient data is to reset entire disk that deletes all
the archived patient data in the disk.
It is strongly recommended that all the patient data be stored in PACS or backup to external media.
Figure 3-18 Disk Encryption - Change key