EasyManuals Logo

GE LOGIQ E10 Service Manual

GE LOGIQ E10
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #231 background imageLoading...
Page #231 background image
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 25
SSA License
SSA License provides a way to do the following:
• When inserted, view the details of an SSA key.
• View the status of the service class options.
• Restore an SSA license when the SSA key is not validating or when a remote log in shows as a
Class A user.
The information on SSA License is available to all service class licenses.
To access SSA License, select Utility (second page) > Service > Utilities > SSA License.
Figure 7-15 SSA License
This table shows all the elements available on SSA License with descriptions.
Table 7-7 SSA License
Element DESCRIPTION
Class M Key Details
Class M Key Status
Status of the SSA key. Valid values are:
• Not Plugged In
• Plugged In
Drive Letter Drive where the SSA key is plugged into the LOGIQ E10.
Expire Date Date the SSA key is set to expire.
SSO ID Identifier for the user assigned to the SSA key.
Key Counter Value Number of times the SSA key has been used.
Max Key Counter Value Number of remaining times the SSA key can be used.
Service Option Keys
Name Name of the service class option.
Status
Status of the access to the associated service class option. Valid values are:
• True
• False
Restore SSA Restores the SSA license to the SSA key.

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
TouchscreenYes
3D/4D ImagingYes
ApplicationsVascular, Cardiac, Musculoskeletal, OB/GYN
Image StorageUSB, Network
ConnectivityDICOM, Ethernet, Wireless