DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
7 - 26 Section 7-3 - Service desktop
To view the Class M license information:
1.) Insert the SSA key.
2.) Navigate to Utility (second page) > Service > Utilities > SSA License.
3.) Under Class M key Details, view the values. For example, the SSO ID for the user assigned to the
SSA key.
If Not Available displays for all of these values, the SSA key is not validating.
To restore an SSA key that is not validating:
1.) Remove the SSA key from the LOGIQ E10.
2.) If open, close the Service desktop.
3.) Restart the LOGIQ E10.
4.) Once the LOGIQ E10 has restarted, plug in the SSA key.
5.) Navigate to Utility (second page) > Service > Utilities > SSA License.
6.) Click Restore SSA.
7.) Check to see if the SSA key validates.
For more information, see:
• "Utilities" on page 7-19