DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 10 Section 8-4 - Loading / Reloading / Upgrading the Software
8-4-4 Space management - moving all images
Refer to the latest revision of the LOGIQ E10 Basic User Manual, Chapter 9 Recording Images.
Speak with the personnel at the site to determine which patient images need to be backed up prior to
starting.
NOTICE
In order to complete a successful restore of the Patient Database, if needed, the images must be
moved away from LOGIQ E10 before doing backup of the Patient Database. Depending on the location
set-up, either move the images to a remote server or to removable media like DVD or CD discs.
As the images are moved, the database will point to the new location.
If the backup procedure is not completed correctly, the images and database information could be lost.
CAUTION
Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
TouchscreenYes
3D/4D ImagingYes
ApplicationsVascular, Cardiac, Musculoskeletal, OB/GYN
Image StorageUSB, Network
ConnectivityDICOM, Ethernet, Wireless

Related product manuals