DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 18 Section 8-4 - Loading / Reloading / Upgrading the Software
Loading C Partition Only Software
CAUTION
Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
Table 8-9 Loading C Partition Only Software
Steps Corresponding Graphic
1.
With the system off, insert the Base Image
and Application Software Installation
Thumb Drive into the USB port on the back
of the LOGIQ E10.
Turn ON the system.
2.
Press any key to confirm you want to load
the software.
3.
Type C_Only. To load C Partition
Press the Enter key.
4.
Let the Load process continue. It will take
several minutes. DO NOT disturb the
process.