DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 25
Loading from Internal Partition Software
A Service Key is not required to run this procedure, if run remotely.
NOTE: LOTO is not required during this procedure.
Use this procedure to reload Base Load and Application Software from the repository partition in the
harddrive.
Table 8-14 Manpower / Time and Tools
Manpower /
Time
Tools
One person /
1 Hour
No tools required.
CAUTION
Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
Table 8-15 Loading from Internal Partition Software
Steps Corresponding Graphic
1.
Log into Service Desktop (if FE is reloading locally, SSA key is required. IF Reload is triggered
remotely by the OLC, no SSA key is required)
Enter Utility tab and select Disruptive mode to enable it
Under Utility tab select software reload
2.
Before proceeding you need to check for
encryption. System asks to confirm if
encryption key is available, if confirm select
second bullet.
If Agree, check box and hit continue reload
3.
System message will appear on the screen
stating that Reload is configured and
system will automatically restart in 10
seconds.
4.
Wait until the system goes through the
whole loading process. System will boot up
to applications automatically after about 30
minutes.