DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 101
Monitor Arm Assembly Adapter removal
Table 8-112 Arm Adapter removal
Steps Corresponding Graphic
1.
Remove the Bulkhead Cover.
2.
Disconnect the Power Cable (P1), the
HDMI (P10), the USB (P3) and the Ground
Cable (circled) from the Adapter ONLY.
The grounding screw is located inside the
adapter.
3.
Remove the Set Screw using a 5 mm hex
wrench.
Carefully lift the Arm Assembly up from the
LOGIQ E10.