DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 334 Section 8-11 - Front End Acquisition / Card Cage parts replacement
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Connect cables and Probes removed earlier.
2.) Power up the system to verify that it operates as intended.
NOTE: Diagnostics can be performed on a Battery when the battery condition is over 15%. The Battery
FRU is shipped at 20~30%. You do not need to check the capacity as a functional test. If the
capacity is less than 15% (normally this does not happen), you will have to wait for it to charge,
it takes around 5 minutes to increase 1%.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-343 Batteries replacement Functional Checks
See:
Section
Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, (latest revision). Leakage
Current measured at (record the value) and
meets allowable limits. Equipment passed all
required checks and is ready for use.
4-2-7 Probe/Connectors Checks
10-8-3 Grounding continuity
10-8-4 Chassis Leakage Current Test
Verify Power Assistant Option functions,
if present.
4-2-4 Power SHUT DOWN