EasyManua.ls Logo

GE LOGIQ E10

GE LOGIQ E10
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 10 Care & Maintenance 10 - 19
10-7-6 Data sheet for enclosure/chassis leakage current
Table 10-13 on page 10-19 shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
Table 10-13 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions Measurement/Test Point Location
System
Power Grounding/PE
Rear
Panel
Lower
Frame
Probe
Connector
Main
Handle
Off Closed
Off Open
On Closed
On Open

Table of Contents

Other manuals for GE LOGIQ E10

Related product manuals