DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 10 Care & Maintenance 10 - 19
10-7-6 Data sheet for enclosure/chassis leakage current
Table 10-13 on page 10-19 shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
Table 10-13 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions Measurement/Test Point Location
System
Power Grounding/PE
Rear
Panel
Lower
Frame
Probe
Connector
Main
Handle
Off Closed
Off Open
On Closed
On Open