DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
2 - 6 Section 2-2 - General LOGIQ E10 requirements
2-2-3 EMI limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. Ultrasound machines also generate EMI. The
LOGIQ E10 complies with limits as stated on the EMC label. However there is no guarantee that
interference will not occur in a particular installation.
Possible EMI sources should be identified before the LOGIQ E10 is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. These
sources include:
medical lasers
scanners
cauterizing guns
•computers
•monitors
fans
gel warmers
microwave ovens
light dimmers
mobile phones
in-house wireless phones (DECT phones)
wireless computer keyboard and mouse
air conditioning system
High Frequency (HF) surgery equipment
general AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.