DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
2 - 8 Section 2-2 - General LOGIQ E10 requirements
2-2-4 Probe environmental requirements
Probes should be operated, stored, and transported according to the listed environmental requirements.
For specific information on environmental requirements for the probes, see “Environmental
Requirements” in chapter 11 of the Basic User Manual.
NOTE: LOGIQ E10 environmental specifications may be different than the probes.
2-2-5 Time and manpower requirements
Site preparation takes time. Begin pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
NOTICE
STORAGE TEMPERATURES VARY DEPENDING ON THE TYPE OF PROBE. WHEN EXPOSED TO
LARGE TEMPERATURE VARIATIONS, THE LOGIQ E10 AND PROBES SHOULD BE KEPT AT
ROOM TEMPERATURE BEFORE USE.
Refer to the table in "If the LOGIQ E10 is very cold or hot" on page 2-2 to determine the needed
settlement time.
CAUTION
RISK OF INJURY OR DAMAGE TO EQUIPMENT
HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE LOGIQ
E10. ATTEMPTS BY ONE PERSON TO MOVE THE LOGIQ E10
CONSIDERABLE DISTANCES, OR ON AN INCLINE, COULD RESULT IN
INJURY, DAMAGE, OR BOTH.

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
TouchscreenYes
3D/4D ImagingYes
ApplicationsVascular, Cardiac, Musculoskeletal, OB/GYN
Image StorageUSB, Network
ConnectivityDICOM, Ethernet, Wireless

Related product manuals