5.1.1 Select the tray and Cellbag bioreactor
Select the Cellbag bioreactor size and corresponding tray according to application re-
quirements and system configuration. See the table below for guidelines.
TrayCellbag bioreactor
size (L)
Culture volume/
bioreactor (L)
Dual modeSingle mode
Tray 20Tray 10, Tray 2020.3 to 1
Tray 20Tray 10, Tray 20100.5 to 5
N/ATray 20201 to 10
Tray 50Tray 50221 to 10
N/ATray 50505 to 25
To use a 20 L or 50 L Cellbag bioreactor, the system must be configured for
single mode.
Note:
Depending on application and configuration it might be possible to cultivate
below the recommended minimum volume. However, it is highly recommended
to stay above this volume for applications that require high agitation, and pH
and DO control. The temperature, pH, and DO sensors need to be submerged
in liquid throughout the complete rocking cycle to function correctly.
Note:
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 105
5 Operation
5.1 Set up the system
5.1.1 Select the tray and Cellbag bioreactor

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ReadyToProcess WAVE 25 and is the answer not in the manual?

GE ReadyToProcess WAVE 25 Specifications

General IconGeneral
BrandGE
ModelReadyToProcess WAVE 25
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals