7.4 ReadyToProcess Pump 25
Corrective actionPossible causeError symptom
Check any warning message and follow the instruc-
tion. If problem persists, contact GE service person-
nel.
An internal error has
occurred, but the pump
is still operating.
Status LED on the
pump rear panel flash-
es red.
Check any warning message and follow the instruc-
tion. If problem persists, contact GE service person-
nel.
An internal error has
occurred, and the
pump is not operating.
Status LED on the
pump rear panel shows
a steady red light.
Contact GE service personnel.An internal error has
occurred, and the
pump is not operating
properly.
Note:
This is normal the first
seconds during power
up.
CAN indicator LED
flashes.
162 ReadyToProcess WAVE 25 Operating Instructions 29009597 AD