7.2 ReadyToProcess WAVE 25 rocker
Rocking
Corrective actionPossible causeError symptom
Try to restart the system a few times.Faulty safety fuses.Rocker does not initial-
ize properly.
If the system still does not initialize properly, shut
it down and contact GE service personnel.
None. This is normal.It takes a while for the
rocker to initialize after
turning on the power.
When starting the run,
the rocker moves too
slowly during the first
cycle.
Check the current alarms.
Reset power to the unit.
If you still get motor alarms, contact GE service
personnel.
Rocker is in an error
state.
Rocker is not rocking.
Make sure that the tray is in normal position.
If this does not help, contact GE service person-
nel.
Tray is in tilt position.
Locate and remove the restricting object.Rocker is mechanically
restricted from moving.
Rocker stops rocking.
Incorrect rocking angle.
This is normal when DO control is enabled. If not
required, disable DO control from the process pic-
ture.
DO control using speed
is enabled. If the DO
deviates from the
setpoint, the rocking
speed changes.
Rocking speed varies.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 157
7 Troubleshooting
7.2 ReadyToProcess WAVE 25 rocker

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ReadyToProcess WAVE 25 and is the answer not in the manual?

GE ReadyToProcess WAVE 25 Specifications

General IconGeneral
BrandGE
ModelReadyToProcess WAVE 25
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals