6.2 Cleaning
Cleaning procedure
To prevent microbial or cross contamination, ReadyToProcess WAVE 25 should be cleaned
after each cultivation. The system must be turned off and unplugged before cleaning.
Clean the exterior of the system units with a damp cloth and a suitable cleaning
agent.
Make sure to clean the temperature sensor arms on the underside of the rocker
platform. If dirt accumulates in these arms, they may not function properly, which
could cause incorrect temperature regulation.
Recommended cleaning agents
All system units can be cleaned with ethanol, isopropanol and Virkon™ at suitable con-
centrations. Refer to Section 8.1 Chemical resistance, on page 166.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 153
6 Maintenance
6.2 Cleaning