5.4.3 End a run
End cultivation and harvest the
culture
Follow the instructions below to end the run and harvest the culture.
ActionStep
Prepare the harvest vessel.1
Click the stop button in the toolbar in System Control to stop the run.2
When asked if you want to end the run, click OK.
Note:
By default, the tray will prepare for tilt at the end of the run. This setting can
be changed in System Settings:Rocker:Prepare for tilt at END.
Hold the textured grip area on each side of the tray, and in one movement,
pull the tray upwards and against you to position the tray into tilt position.
3
Connect tubing from the Cellbag bioreactor to the harvest vessel.4
Empty the Cellbag bioreactor using a pump or gravity flow.5
Disconnect the tubing from the Cellbag bioreactor to the harvest vessel.6
Procedures after harvest
Follow the instructions below when the cultivation is finished and the culture is harvested.
ActionStep
Clamp off the inlet and outlet vent filters of the Cellbag bioreactor.1
Disconnect the tubing from the inlet vent filter on the Cellbag bioreactor.2
Disconnect any other tubing and cables still connected to the Cellbag
bioreactor.
3
Release and remove the empty Cellbag bioreactor from the tray by pressing
down the bag clamp opener.
4
148 ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
5 Operation
5.4 Perform cultivation
5.4.3 End a run