Corrective actionPossible causeError symptom
Verify the connection between the rocker and the
client computer or network.
Rocker does not have
any connection to the
UNICORN database.
Power button flashes
red.
Disable the missing components in UNICORN. See
Section 5.2.3 Configure system properties, on
page 127.
System units such as
CBCU or pump defined
in the system setup in
UNICORN are not con-
nected to the rocker.
Connect the missing components to the rocker.
See Connect system units to the rocker, on page 86.
Read any warning message carefully and follow
the instructions. If problem persists, contact GE
service personnel.
An internal error has
occured in the embed-
ded computer of the
rocker.
•
Wait one minute. If the system still does not
shut down, keep pressing the power button to
force system shut down.
•
If problem persists, contact GE service person-
nel.
The embedded comput-
er of the rocker fails to
shut down.
System does not shut
down when the power
button is pressed.
None. This is normal.At the first command
sent to the system, the
system enters Run
state and the rocker is
initialized and makes a
few movements.
Rocking starts when
sending another com-
mand than Start rock-
ing.
158 ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
7 Troubleshooting
7.2 ReadyToProcess WAVE 25 rocker